Trigger.blender[N].Event_Id; Trigger.blender[N].Orenable - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Example
trigger.blender[2].clear()
Also see
None

trigger.blender[N].EVENT_ID

This constant contains the trigger blender event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
N
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to trigger
events from this trigger blender.
Example
digio.trigger[1].stimulus = trigger.blender[2].EVENT_ID
Also see
None

trigger.blender[N].orenable

This attribute selects whether the blender operates in OR mode or AND mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
N
2600BS-901-01 Rev. B / May 2013
Affected by
Trigger event number
The blender number (1 to 6)
Affected by
Instrument reset
Trigger blender N reset
Recall setup
The orenable mode:
true: OR mode
false: AND mode
The trigger blender (1 to 6)
Section 7: TSP command reference
Clears the event detector for
blender 2.
Where saved
Default value
Set the trigger stimulus of
digital I/O trigger 1 to be
controlled by the trigger
blender 2 event.
Where saved
Default value
Not saved
false (AND mode)
7-337

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents