Series 2600B System SourceMeter® Instrument Reference Manual

smuX.sense

This attribute contains the state of the sense mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
senseMode = smuX.sense
smuX.sense = senseMode
senseMode
X
Details
Source-measure operations are performed using either 2-wire local sense connections or 4-wire remote sense
connections. Writing to this attribute selects the sense mode.
The smuX.SENSE_CALA mode is only used for calibration and may only be selected when calibration is enabled.
The sense mode can be changed between local and remote while the output is on.
The calibration sense mode cannot be selected while the output is on.
Resetting the instrument selects the local sense mode.
Example
smua.sense = smua.SENSE_REMOTE
Also see
2-wire local sensing connections
4-wire remote sensing connections
Sense mode selection

smuX.source.autorangeY

This attribute contains the state of the source autorange control (on/off).
Type
TSP-Link accessible
Attribute (RW)
Yes
2600BS-901-01 Rev. B / May 2013
Affected by
SMU reset
Instrument reset
Recall setup
The sense mode; set to one of the following:
0 or smuX.SENSE_LOCAL: Selects local sense (2-wire)
1 or smuX.SENSE_REMOTE: Selects remote sense (4-wire)
3 or smuX.SENSE_CALA: Selects calibration sense mode
Source-measure unit (SMU) channel (for example, smua.sense applies to SMU
channel A)
(on page 2-53)
(on page 2-54)
(on page 2-76)
Affected by
SMU reset
Instrument reset
Recall setup
Section 7: TSP command reference
Where saved
Default value
Saved setup
0 (smuX.SENSE_LOCAL)
Selects remote sensing for SMU
channel A.
Where saved
Default value
Saved setup
1 (smuX.AUTORANGE_ON)
7-217