Section 7: TSP command reference
Details
This setting controls the aperture of measurements made for contact check. It does not affect the
smuX.measure.nplc aperture setting.
The speed setting can have a dramatic effect on the accuracy of the measurement (see specifications).
Example
smua.contact.speed = smua.CONTACT_SLOW
Also see
Contact check connections
Contact check measurements
reset()
(on page 7-160)
smuX.contact.check()
smuX.contact.r()
smuX.reset()
(on page 7-216)

smuX.contact.threshold

This attribute stores the resistance threshold for the smuX.contact.check() function. This attribute is not
available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
rValue = smuX.contact.threshold
smuX.contact.threshold = rValue
rValue
X
Details
The threshold should be set to less than 1 kΩ.
Example
smua.contact.threshold = 5
Also see
Contact check connections
Contact check measurements
reset()
(on page 7-160)
smuX.contact.check()
smuX.reset()
(on page 7-216)
7-196
(on page 2-54)
(on page 2-44)
(on page 7-193)
(on page 7-194)
Affected by
Instrument reset
SMU reset
Recall setup
The resistance above which contact check should fail (measured in ohms)
Source-measure unit (SMU) channel (for example, smua.contact.threshold
applies to SMU channel A)
(on page 2-54)
(on page 2-44)
(on page 7-193)
Series 2600B System SourceMeter® Instrument Reference Manual
Configure contact check for higher accuracy
on SMU channel A.
Where saved
Not saved
Set the contact check threshold for SMU channel A to 5 Ω.
Default value
50 (50 Ω)