Smux.source.limity - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Details
This attribute configures the output level of the voltage or current source.
If the source is configured as a voltage source and the output is on, the new smuX.source.levelv setting is
sourced immediately. If the output is off or the source is configured as a current source, the voltage level is
sourced when the source is configured as a voltage source and the output is turned on.
If the source is configured as a current source and the output is on, the new smuX.source.leveli setting is
sourced immediately. If the output is off or the source is configured as a voltage source, the current level is
sourced when the source is configured as a current source and the output is turned on.
The sign of sourceLevel dictates the polarity of the source. Positive values generate positive voltage or current
from the high terminal of the source relative to the low terminal. Negative values generate negative voltage or
current from the high terminal of the source relative to the low terminal.
The reset() function sets the source levels to 0 V and 0 A.
Example
smua.source.levelv = 1
Also see
smuX.source.compliance
smuX.source.func
smuX.source.output
Source-measure concepts

smuX.source.limitY

This attribute sets compliance limits.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
limit = smuX.source.limitY
smuX.source.limitY = limit
limit
X
Y
2600BS-901-01 Rev. B / May 2013
(on page 7-219)
(on page 7-221)
(on page 7-228)
(on page 4-1)
Affected by
SMU reset
Instrument reset
Recall setup
The compliance limit value; set to one of the following values:
Voltage compliance:
Models 2601B/2602B/2604B: 10 mV to 40 V
Models 2611B/2612B/2614B/2634B/2635B/2636B: 20 mV to 200 V
Current compliance:
Models 2601B/2602B/2604B/2611B/2612B/2614B: 10 nA to 3 A
Models 2634B/2635B/2636B: 100 pA to 1.5 A
Power compliance (in watts)
Source-measure unit (SMU) channel (for example, smua.source.limitv applies to SMU
channel A)
SMU function (v = voltage, i = current)
Section 7: TSP command reference
Sets voltage source of SMU
channel A to 1 V.
Where saved
Default value
Saved setup
Limit voltage:
Models 2601B/2602B/2604B: 40 (40 V)
Models
2611B/2612B/2614B/2634B/2635B/2636B:
20 (20 V)
Limit current:
Models 2601B/2602B/2604B: 1 (1 A)
Models
2611B/2612B/2614B/2634B/2635B/2636B:
100e-3 (100 mA)
Limit power: 0 (disabled)