Usb Storage Overview - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 2: General operation
To set the current limit in NORMAL output-off mode remotely:
smuX.source.offlimiti = iValue
To set the voltage limit in NORMAL output-off mode remotely:
smuX.source.offlimitv = vValue
* smuX: For Models 2601B, 2611B, and 2635B, this value is smua (SMU Channel A); for Models
2602B, 2604B, 2612B, 2614B, 2634B, and 2636B, this value can be smua (for SMU Channel A) or
smub (for SMU Channel B).
Remote programming output-off states quick reference
The content of the following table is a quick reference of commands for programming output-off states
from a remote interface.
Output-off state programming quick reference
Command*
smuX.source.offmode = smua.OUTPUT_NORMAL
smuX.source.offmode = smua.OUTPUT_HIGH_Z
smuX.source.offmode = smua.OUTPUT_ZERO
smuX.source.offfunc = smua.OUTPUT_DCVOLTS
smuX.source.offfunc = smua.OUTPUT_DCAMPS
smuX.source.offlimiti = iValue
smuX.source.offlimitv = vValue
* smuX: For Models 2601B, 2611B, and 2635B, this value is smua (SMU Channel A); for Models 2602B,
2604B, 2612B, 2614B, 2634B, and 2636B, this value can be smua (for SMU Channel A) or smub (for SMU
Channel B).

USB storage overview

The Series 2600B System SourceMeter
scripts and to transfer files from the instrument to the host computer, you need a USB flash drive.
For information about saving reading buffers to the USB flash drive, see
(on page 3-9).
For information about storing and loading scripts to and from the USB flash drive, see
user script
For information about file I/O, see
For information about saving user setups, see
2-80
(on page 6-8).
File I/O
Series 2600B System SourceMeter® Instrument Reference Manual
Description
Selects normal output-off mode.
Selects high-impedance output-off
mode.
Selects zero output-off mode.
Sets 0 V output with current limit
specified by the
smua.source.offlimiti
attribute.
Sets 0 A output with voltage limit
specified by the
smua.source.offlimitv
attribute.
Sets current limit in normal
output-off mode.
Sets voltage limit in normal
output-off mode.
®
instrument includes a USB port on the front panel. To store
(on page 5-6).
Saved setups
Saving reading buffers
(on page 2-46).
2600BS-901-01 Rev. B / May 2013
Save a