Series 2600B System SourceMeter® Instrument Reference Manual

ConfigPulseVMeasureISweepLog()

This
KIPulse factory script
Type
TSP-Link accessible
Function
No
Usage
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout, sync_in_abort)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in)
f, msg = ConfigPulseVMeasureISweepLog(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag)
f
msg
smu
bias
start
stop
limit
ton
toff
points
buffer
tag
sync_in
sync_out
sync_in_timeout
sync_in_abort
1
The KIPulse factory script provides examples of how to generate pulses and to provide a simple pulsing
interface. Pulses can be generated using the functions listed below.@Please note the following information about
the KIPulse factory script:This factory script only operates on the channels present in the instrument executing the
pulse functions. These functions will not operate correctly if you attempt to access instrument channels over the
TSPLink® interface.The KIPulse factory scripts are ge ...
2600BS-901-01 Rev. B / May 2013
1
function configures a voltage pulse train with a current measurement at each point.
Affected by
A Boolean flag; this flag is true when the pulse was successfully configured,
false when errors were encountered
A string message; if the f flag is false, msg contains an error message; if it is
true, msg contains a string indicating successful configuration
System SourceMeter
channel A)
Bias level in volts
Pulse sweep start level in volts
Pulse sweep stop level in volts
Current limit (for example, compliance) in amperes
Pulse on time in seconds
Pulse off time in seconds
Number of pulse-measure cycles
Reading buffer where pulsed measurements will be stored; if this is nil when the
function is called, no measurements will be made when the pulse train is initiated
Numeric identifier to be assigned to the defined pulse train
Defines a digital I/O trigger input line; if programmed, the pulse train will wait for a
trigger input before executing each pulse
Defines a digital I/O trigger output line; if programmed, the pulse train will generate a
trigger output immediately before the start of ton
Specifies the length of time (in seconds) to wait for input trigger; default value is 10 s
Specifies whether or not to abort pulse if input trigger is not received; if pulse aborts
because of a missed trigger, a timer timeout message is returned; true or false
Where saved
Default value
®
instrument channel (for example, smua refers to SMU
Section 7: TSP command reference
7-43