Smux.source.func; Smux.source.highc - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Example
smua.source.delay = smua.DELAY_AUTO
Also see
reset()
(on page 7-160)
smuX.measure.count
smuX.measure.delay
smuX.reset()

smuX.source.func

This attribute contains the source function, which can be voltage or current.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sFunction = smuX.source.func
smuX.source.func = sFunction
sFunction
X
Details
Reading this attribute indicates the output function of the source for the specified SMU channel. Setting this
attribute configures the specified SMU channel as either a voltage source or a current source.
Example
smua.source.func = smua.OUTPUT_DCAMPS
Also see
smuX.source.levelY
smuX.source.output

smuX.source.highc

This attribute enables or disables high-capacitance mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
2600BS-901-01 Rev. B / May 2013
(on page 7-201)
(on page 7-202)
(on page 7-216)
Affected by
SMU reset
Instrument reset
Recall setup
The source function; set to one of the following values:
0 or smuX.OUTPUT_DCAMPS: Selects the current source function
1 or smuX.OUTPUT_DCVOLTS: Selects the voltage source function
Source-measure unit (SMU) channel (for example, smua.source.func applies to
SMU channel A)
(on page 7-222)
(on page 7-228)
Affected by
SMU reset
Instrument reset
Recall setup
Section 7: TSP command reference
Sets the delay for SMU channel A
to automatic (a range-dependent
delay is inserted whenever the
source is changed).
Where saved
Default value
Saved setup
1 (smuX.OUTPUT_DCVOLTS)
Sets the source function of SMU
channel A to be a current source.
Where saved
Default value
Saved setup
0 (smuX.DISABLE)
7-221

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all