Status.questionable.instrument - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
As an example, to set bit B1 of the questionable status calibration summary enable register, set
status.questionable.calibration.enable = status.questionable.calibration.SMUA.
In addition to the above constants, questionableRegister can be set to the numeric equivalent of the bit to
set. To set more than one bit of the register, set operationRegister to the sum of their decimal weights. For
example, to set bits B1 and B2, set questionableRegister to 6 (which is the sum of 2 + 4).
Example
status.questionable.calibration.enable =
status.questionable.calibration.SMUA
Also see
Questionable Status Registers
status.questionable.*

status.questionable.instrument.*

This attribute contains the questionable status instrument summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
questionableRegister = status.questionable.instrument.condition
questionableRegister = status.questionable.instrument.enable
questionableRegister = status.questionable.instrument.event
questionableRegister = status.questionable.instrument.ntr
questionableRegister = status.questionable.instrument.ptr
status.questionable.instrument.enable = questionableRegister
status.questionable.instrument.ntr = questionableRegister
status.questionable.instrument.ptr = questionableRegister
questionableRegister
2600BS-901-01 Rev. B / May 2013
(on page E-13)
(on page 7-304)
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The status of the questionable status instrument summary register; a
zero (0) indicates no bits set (also send 0 to clear all bits); other values
indicate various bit settings
Section 7: TSP command reference
Uses a constant to set the SMUA bit
of the questionable status calibration
summary enable register.
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
(All bits set)
7-307