Status.measurement.buffer_Available - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference

status.measurement.buffer_available.*

This attribute contains the measurement event buffer available summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
measurementRegister = status.measurement.buffer_available.condition
measurementRegister = status.measurement.buffer_available.enable
measurementRegister = status.measurement.buffer_available.event
measurementRegister = status.measurement.buffer_available.ntr
measurementRegister = status.measurement.buffer_available.ptr
status.measurement.buffer_available.enable = measurementRegister
status.measurement.buffer_available.ntr = measurementRegister
status.measurement.buffer_available.ptr = measurementRegister
measurementRegister
Details
These attributes are used to read or write to the measurement event buffer available summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For
example, assume value 6 is returned for the enable register. The binary equivalent is 0000 0000 0000 0110. This
value indicates that bit B1 (SMUA) and bit B2 (SMUB) are set.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
page E-1) and
the following table.
Bit
Value
B0
Not used
B1
status.measurement.buffer_available.SMUA
B2
status.measurement.buffer_available.SMUB
B3-B15
Not used
7-258
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The status of the measurement event register; a zero (0) indicates no bits set
(also send 0 to clear all bits); other values indicate various bit settings
Enable and transition registers
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B:
6 (All bits set)
(on page E-19). The individual bits of this register are defined in
Description
Not applicable.
Set bit indicates that there is at least one
reading stored in either or both of the dedicated
reading buffers.
Bit B1 decimal value: 2
Binary value: 0000 0010
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates that there is at least one reading
stored in either or both of the dedicated reading
buffers.
Bit B2 decimal value: 4
Binary value: 0000 0100
Not applicable.
Status register set contents
2600BS-901-01 Rev. B / May 2013
(on

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents