setup.save()

This function saves the present setup as a user-saved setup.
Type
TSP-Link accessible
Function
Yes
Usage
setup.save(id)
id
Details
When the id parameter is an integer (n), it is interpreted as the setup number to save to the instrument's
nonvolatile memory.
When you save to a specified integer (1 to 5) in nonvolatile memory, the previous setup at that same
location is overwritten.
When the id parameter is a string, it is interpreted as the path and file name of the location to save the present
setup on a USB flash drive. The path may be absolute or relative to the current working directory.
Example
setup.save(5)
Also see
Saved setups
setup.recall()

smuX.abort()

This function terminates all overlapped operations on the specified source-measure unit (SMU).
Type
TSP-Link accessible
Function
Yes
Usage
smuX.abort()
X
Details
The smuX.abort() function does not turn the output off or change any settings.
If this function is used to abort a sweep, when it is executed, the SMU exits its trigger model immediately and
returns to the trigger model's idle state.
2600BS-901-01 Rev. B / May 2013
Affected by
An integer or string specifying where to save the user setup:
Save in nonvolatile memory (1 to 5)
Save as user-saved setup on a USB flash drive ("/path/filename")
(on page 2-46)
(on page 7-180)
Affected by
Source-measure unit (SMU) channel (for example, smua.abort() applies to SMU
channel A)
Section 7: TSP command reference
Where saved
Default value
Saves the present setup to the internal memory of
the instrument at location 5.
Where saved
Default value
7-181