Keithley 2601B Reference Manual page 382

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference
This function does not cause the specified smu to output a pulse train. It does check to see if all the pulse
dimensions can be achieved, and if they can, assigns the indicated tag or index to the pulse train.
The InitiatePulseTest(tag) and InitiatePulseTestDual(tag1, tag2) functions are used to initiate
a pulse train assigned to a valid tag.
Example
ConfigPulseVMeasureISweepLin(smua, 0, 1,
10, 1, 10e-3, 20e-3, 16,
smua.nvbuffer1, 4)
Also see
InitiatePulseTest()
InitiatePulseTestDual()
KIPulse factory script
7-42
Series 2600B System SourceMeter® Instrument Reference Manual
Figure 131: ConfigPulseVMeasureISweepLin()
(on page 7-103)
(on page 7-104)
(on page 5-21)
Set up a pulsed sweep that uses System
®
SourceMeter
instrument channel A. The pulsed
sweep starts at 1 V, ends at 10 V, and returns to a
0 V bias level between pulses. Each pulsed step is on
for 10 ms, and then at the bias level for 20 ms.
The current limit is 1 A during the entire pulsed
sweep. The pulse train is comprised of 16 pulsed
steps, and the pulse train is assigned a tag index of
4.
2600BS-901-01 Rev. B / May 2013

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents