Series 2600B System SourceMeter® Instrument Reference Manual
Usage
PulseIMeasureV(smu, bias, level, ton, toff, points)
smu
bias
level
ton
toff
points
Details
Data for pulsed voltage measurements, current levels, and timestamps are stored in smua.nvbuffer1.
If any parameters are omitted or nil, the operator is prompted to enter them using the front panel.
To perform the specified number of pulse I, measure V cycles, this function:
1.
Sets the smu to output bias amperes and dwell for toff seconds.
2.
Sets the smu to output level amperes and dwell for ton seconds.
3.
Performs voltage measurement with source at level amperes.
4.
Sets the smu to output bias amperes for toff seconds.
5.
Repeats steps 2 through 4 for all remaining points pulse-measure cycles.
Example
PulseIMeasureV(smua, 0.001, 1.0,
20e-3, 40e-3, 10)
Also see
KIPulse factory script

PulseVMeasureI()

This KIPulse factory script function performs a specified number of pulse V, measure I cycles.
Type
TSP-Link accessible
Function
No
2600BS-901-01 Rev. B / May 2013
®
System SourceMeter
instrument channel (for example, smua refers to SMU
channel A)
Bias level in amperes
Pulse level in amperes
Pulse on time in seconds
Pulse off time in seconds
Number of pulse-measure cycles
Figure 134: PulseIMeasureV
(on page 5-21)
Affected by
Section 7: TSP command reference
SMU A outputs 1 mA and dwells for 40 ms, outputs
1 A and dwells for 20 ms. The voltage measurements
occur during each 20 ms dwell period. After the
measurement, the output returns to 1 mA and dwells
for 40 ms. This pulse-measure process repeats nine
more times.