Status.operation.instrument.digio.trigger_Overrun.* - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference
Details
These attributes are used to read or write to the operation status digital I/O summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
page E-1) and
the following table.
Bit
Value
B0-B9
Not used
B10
status.operation.instrument.digio.TRIGGER_OVERRUN
status.operation.instrument.digio.TRGOVR
B11-B15
Not used
In addition to the above constant, operationRegister can be set to the decimal equivalent of the bit to set.
Example
status.operation.instrument.digio.enable =
status.operation.instrument.digio.TRGOVR
Also see
Operation Status Registers

status.operation.instrument.digio.trigger_overrun.*

status.operation.instrument.digio.trigger_overrun.*
This attribute contains the operation status digital I/O overrun register set. This attribute is not available on the
Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
7-276
Enable and transition registers
(on page E-9)
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
Series 2600B System SourceMeter® Instrument Reference Manual
(on page E-19). The individual bits of this register are defined in
(on page 7-276)
Where saved
- -
Not saved
Not saved
Not saved
Not saved
Not saved
Status register set contents
Description
Not applicable
Set bit indicates an
enabled bit in the
Operation Status Digital
I/O Overrun Register is set.
Bit B10 decimal value:
1,024
Binary value:
0100 0000 0010
Not applicable
Sets the TRGOVR bit of the
operation status digital I/O summary
enable register using a constant.
Default value
- -
Not applicable
0
0
0
32,766 (All bits set)