Section 2: General operation
Voltage and current
The following table lists the source and measure limits for the voltage and current functions. The full
range of operation is explained in
Source-measure capabilities
Model 2601B/2602B
Range
Source
100 mV
±101 mV
1 V
±1.01 V
6 V
±6.06 V
40 V
±40.4 V
100 nA
±101 nA
1 µA
±1.01 µA
10 µA
±10.1 µA
100 µA
±101 µA
1 mA
±1.01 mA
10 mA
±10.1 mA
100 mA
±101 mA
1 A
±1.01 A
3 A
±3.03 A
Max Power = 40.4 W per channel
Source-measure capabilities
Model 2601B/2602B/2604B
Range
Source
100 mV
±101 mV
1 V
±1.01 V
6 V
±6.06 V
40 V
±40.4 V
100 nA
±101 nA
1 µA
±1.01 µA
10 µA
±10.1 µA
100 µA
±101 µA
1 mA
±1.01 mA
10 mA
±10.1 mA
100 mA
±101 mA
1 A
±1.01 A
3 A
±3.03 A
2-26
Series 2600B System SourceMeter® Instrument Reference Manual
Operating boundaries
Model 2611B/2612B
Measure
Range
Source
±102 mV
200 mV
±202 mV
±1.02 V
2 V
±2.02 V
±6.12 V
20 V
±20.2 V
1
±40.8 V
±202 V
200 V
±102 nA
±101 nA
100 nA
±1.02 µA
±1.01 µA
1 µA
±10.2 µA
±10.1 µA
10 µA
±102 µA
±101 µA
100 µA
±1.02 mA
±1.01 mA
1 mA
±10.2 mA
±10.1 mA
10 mA
±102 mA
±101 mA
100 mA
±1.02 A
±1.01 A
1 A
±3.06 A
±1.515 A
1.5 A
2
±10.1 A
10 A
Max Power = 30.603 W per channel
1.
200 V source range available only when
interlock is enabled. See Digital I/O (on page
3-83).
2.
10 A range available only in pulse mode.
Model 2611B/2612B/2614B
Measure
Range
Source
±102 mV
200 mV
±202 mV
±1.02 V
2 V
±2.02 V
±6.12 V
20 V
±20.2 V
1
±40.8 V
±202 V
200 V
±102 nA
±101 nA
100 nA
±1.02 µA
±1.01 µA
1 µA
±10.2 µA
±10.1 µA
10 µA
±102 µA
±101 µA
100 µA
±1.02 mA
±1.01 mA
1 mA
±10.2 mA
±10.1 mA
10 mA
±102 mA
±101 mA
100 mA
±1.02 A
±1.01 A
1 A
±3.06 A
±1.515 A
1.5 A
2
±10.1 A
10 A
(on page 4-4).
Model 2635B/2636B
Measure
Range
±204 mV
200 mV
±2.04 V
2 V
±20.4 V
20 V
3
±204 V
200 V
±102 nA
100 pA
±1.02 µA
1 nA
±10.2 µA
10 nA
±102 µA
100 nA
±1.02 mA
1 µA
±10.2 mA
10 µA
±102 mA
100 µA
±1.02 A
1 mA
±1.53 A
10 mA
±10.2 A
100 mA
1 A
1.5 A
Max Power = 30.603 W per channel
3.
200 V source range available only when
interlock is enabled. See Digital I/O (on page 3-
83).
Model 2635B/2636B
Measure
Range
±204 mV
200 mV
±2.04 V
2 V
±20.4 V
20 V
3
±204 V
200 V
±102 nA
100 pA
±1.02 µA
1 nA
±10.2 µA
10 nA
±102 µA
100 nA
±1.02 mA
1 µA
±10.2 mA
10 µA
±102 mA
100 µA
±1.02 A
1 mA
±1.53 A
10 mA
±10.2 A
100 mA
1 A
1.5 A
2600BS-901-01 Rev. B / May 2013
Source
Measure
±202 mV
±204 mV
±2.02 V
±2.04 V
±20.2 V
±20.4 V
±202 V
±204 V
N/A
±102 pA
±1.01 nA
±1.02 nA
±10.1 nA
±10.2 nA
±101 nA
±102 nA
±1.01 µA
±1.02 µA
±10.1 µA
±10.2 µA
±101 µA
±102 µA
±1.01 mA
±1.02 mA
±10.1 mA
±10.2 mA
±101 mA
±102 mA
±1.01 A
±1.02 A
±1.515 A
±1.53 A
Source
Measure
±202 mV
±204 mV
±2.02 V
±2.04 V
±20.2 V
±20.4 V
±202 V
±204 V
N/A
±102 pA
±1.01 nA
±1.02 nA
±10.1 nA
±10.2 nA
±101 nA
±102 nA
±1.01 µA
±1.02 µA
±10.1 µA
±10.2 µA
±101 µA
±102 µA
±1.01 mA
±1.02 mA
±10.1 mA
±10.2 mA
±101 mA
±102 mA
±1.01 A
±1.02 A
±1.515 A
±1.53 A