Section 7: TSP command reference

smuX.reset()

This function turns off the output and resets the commands that begin with smu. to their default settings.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.reset()
X
Details
This function turns off the output and returns the specified SMU to its default settings.
Example
smua.reset()
Also see
reset()
(on page 7-160)

smuX.savebuffer()

This function saves one source-measure unit (SMU) dedicated reading buffer to nonvolatile memory (there are
two dedicated reading buffers for each SMU).
Type
TSP-Link accessible
Function
Yes
Usage
smuX.savebuffer(smuX.nvbufferY)
X
Y
Details
When the instrument is turned off and back on, the dedicated reading buffers are restored from nonvolatile
memory to their last saved values.
Example
smua.savebuffer(smua.nvbuffer1)
Also see
Reading buffers
savebuffer()
(on page 7-161)
smuX.nvbufferY
7-216
Affected by
Source-measure unit (SMU) channel (for example, smua.reset() applies to SMU
channel A)
Affected by
SMU channel (for example, smua.savebuffer(smua.nvbuffer1) applies to
SMU channel A)
SMU dedicated reading buffer (1 or 2)
(on page 3-6)
(on page 7-215)
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Turns off the output and resets SMU
channel A to its default settings.
Where saved
Default value
Default value
Saves buffer 1 (SMU channel A) to
internal memory.
2600BS-901-01 Rev. B / May 2013