This attribute sets the limit (current or voltage) used when the source-measure unit (SMU) is in normal output-off
mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sourceLimit = smuX.source.offlimitY
smuX.source.offlimitY = sourceLimit
sourceLimit
X
Y
Details
Setting the current limit to lower than 1 mA may interfere with operation of the contact check function. See
smuX.contact.check() and smuX.contact.r() for details.
Example
smua.source.offlimiti = 10e-3
Also see
smuX.contact.check()
smuX.contact.r()
smuX.source.offfunc
smuX.source.offmode
smuX.source.offmode
This attribute sets the source output-off mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
7-226
(on page 2-77)
(on page 7-226)
(on page 7-226)
(on page 7-228)
Affected by
SMU reset
Instrument reset
Recall setup
Set to the limit to be used when the SMU is in normal output-off mode
Source-measure unit (SMU) channel (for example, smua.source.offlimiti
applies to SMU channel A)
SMU source function (v = voltage, i = current)
(on page 7-193)
(on page 7-194)
(on page 7-225)
(on page 7-226)
Affected by
SMU reset
Instrument reset
Recall setup
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Saved setup
Where saved
Saved setup
Default value
Current: 1e-3 (1 mA)
Voltage: 40 (40 V)
Changes the normal output-off
mode limit to 10 mA for SMU
channel A.
Default value
0 (smuX.OUTPUT_NORMAL)
2600BS-901-01 Rev. B / May 2013