Appendix B: Calibration

Calibration commands quick reference

The following table summarizes remote calibration commands. For a more complete description of
these commands, refer to the
Calibration commands
Command**
smuX.cal.adjustdate = adjustDate
smuX.cal.date = calDate
smuX.cal.due = calDue
smuX.cal.password = "newPassword"
smuX.cal.polarity = calPolarity
smuX.cal.restore(calset)
smuX.cal.save()
calstate = smuX.cal.state
smuX.cal.unlock("password")
smuX.measure.calibratei(range,
cp1Measured, cp1Reference,
cp2Measured, cp2Reference)
smuX.measure.calibratev(range,
cp1Measured, cp1Reference,
cp2Measured, cp2Reference)
smuX.source.calibratei(range,
cp1Expected, cp1Reference,
cp2Expected, cp2Reference)
smuX.source.calibratev(range,
cp1Expected, cp1Reference,
cp2Expected, cp2Reference)
B-24
Series 2600B System SourceMeter® Instrument Reference Manual
TSP command reference
Description
Set date when the adjustment was done.
Set calibration date (calDate of 0 indicates date not set).
Set date when calibration should be performed (calDue of 0
indicates date not set).
Change password to "newPassword".
Set polarity:
smuX.CAL_AUTO (automatic polarity).
smuX.CAL_NEGATIVE (negative polarity).
smuX.CAL_POSITIVE (positive polarity).
Load set of calibration constants:
smuX.CALSET_NOMINAL (nominal constants).
smuX.CALSET_FACTORY (factory constants).
smuX.CALSET_DEFAULT (normal constants).
smuX.CALSET_PREVIOUS (previous constants).
Store constants in nonvolatile memory as DEFAULT
calibration set.
Request calibration state:
smuX.CALSTATE_CALIBRATING
smuX.CALSTATE_LOCKED
smuX.CALSTATE_UNLOCKED
Unlock calibration (default password: KI0026XX)
Adjust current measurement range calibration*:
±range (measurement range to adjust).
cp1Measured (Series 2600B measured value for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Measured (Series 2600B measured value for cal. point 2).
cp2Reference (reference measurement for cal. point 2).
Adjust voltage measurement range calibration*:
±range (measurement range to adjust).
cp1Measured (Series 2600B measured value for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Measured (Series 2600B measured value for cal. point 2).
cp2Reference (reference measurement for cal. point 2).
Adjust current source range calibration*:
±range (source range to adjust).
cp1Expected (source value programmed for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Expected (source value programmed for cal. point 2).
cp2Reference (reference measurement for cal. point 2).
Adjust voltage source range calibration*:
±range (source range to adjust).
cp1Expected (source value programmed for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Expected (source value programmed for cal. point 2).
cp2Reference (reference measurement for cal. point 2)
(on page 7-1).