Output characteristics:
Configured trigger events, as well as the digio.trigger[N].assert() and
tsplink.trigger[N].assert() commands, cause the physical line state to float high during
the trigger pulse duration.
An action overrun occurs if the physical line state is high while a stimulus event occurs.
Rising edge acceptor trigger mode
The rising edge acceptor trigger mode (RisingA) generates a low pulse and detects rising edge
pulses (see the following figures).
Input characteristics:
All rising edges generate an input event.
2600BS-901-01 Rev. B / May 2013
Figure 74: RisingM output trigger
Figure 75: RisingA input trigger