Status.operation.measuring - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference

status.operation.measuring.*

This attribute contains the operation status measuring summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
operationRegister = status.operation.measuring.condition
operationRegister = status.operation.measuring.enable
operationRegister = status.operation.measuring.event
operationRegister = status.operation.measuring.ntr
operationRegister = status.operation.measuring.ptr
status.operation.measuring.enable = operationRegister
status.operation.measuring.ntr = operationRegister
status.operation.measuring.ptr = operationRegister
operationRegister
Details
These attributes are used to read or write to the operation status measuring summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
page E-1) and
the following table.
Bit
Value
B0
Not used
B1
status.operation.measuring.SMUA
B2
status.operation.measuring.SMUB
B3-B15
Not used
7-296
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The status of the operation status measuring summary register; a zero (0) indicates
no bits set (also send 0 to clear all bits); other values indicate various bit settings
Enable and transition registers
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
Models 2601B/2611B/2635B:
2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B:
6 (All bits set)
(on page E-19). The individual bits of this register are defined in
Description
Not applicable.
Bit will be set when SMU A is taking an
overlapped measurement, but it will not be set
when taking a normal synchronous measurement.
Bit B1 decimal value: 2
Binary value: 0000 0010
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Bit will
be set when SMU B is taking an overlapped
measurement, but it will not be set when taking a
normal synchronous measurement.
Bit B2 decimal value: 4
Binary value: 0000 0100
Not applicable.
Status register set contents
2600BS-901-01 Rev. B / May 2013
(on

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents