Series 2600B System SourceMeter® Instrument Reference Manual
Appendix G: Model 2400 emulation

Model 2400 SCPI command compatibility

The following tables lists all the Model 2400 SCPI commands that are not fully supported by the
Model 2400 personality script (Persona2400) and each command's specific compatibility details.
:CALCulate subsystem
:CALCulate2:NULL:ACQuire
Each time the personality script is run, the null offset value will be reset to zero (0). Until a measurement is taken, this
command will use a zero (0) for the null offset value. After a measurement is taken, this command will use latest
measurement to set the null offset value (the same way as a Model 2400).
:DISPlay subsystem
:DISPlay:DIGits 4|5|6|7|DEFault|MINimum|MAXimum
When 3.5 digits or MINimum is requested, the instrument will set the resolution to 4.5 digits.
:DISPlay:DIGits? [DEFault|MINimum|MAXimum]
When queried for the MINimum, the instrument returns 5.
:DISPlay:ENABle <Bool>
The Series 2600B performance is not degraded by display operation. The instrument will accept the command but no
action is performed.
:DISPlay:WINDow2:ATTRibutes?
The instrument always returns 32 zeros.
:DISPlay:WINDow2:TEXT:STATe <Bool>
Changing this setting also changes the :DISPlay[:WINDow[1]]:TEXT:STATe setting to the same value.
DISPlay[:WINDow[1]]:ATTRibutes?
The instrument always returns 20 zeros.
DISPlay[:WINDow[1]]:TEXT:STATe <Bool>
Changing this setting also changes the :DISPlay:WINDow2:TEXT:STATe setting to the same value.
:ROUte subsystem
:ROUTe:TERMinals FRONt|REAR
The instrument will accept the command and ignore it.
:ROUTe:TERMinals?
The instrument will always return REAR.
2600BS-901-01 Rev. B / May 2013