I_Leakage_Measure(); I_Leakage_Threshold(); Initiatepulsetest() - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Example
smua.source.highc = smua.ENABLE
smua.source.levelv = 5
smua.source.output = smua.OUTPUT_ON
delay(1)
pass = i_leakage_threshold(smua, 0, 1,
300e-3, 10e-6, 100e-3, 1e-6, 1)
Also see
High-capacitance mode

i_leakage_measure()

High-capacitance mode
KIHighC factory script

InitiatePulseTest()

This KIPulse factory script function initiates the pulse configuration assigned to tag.
Type
TSP-Link accessible
Function
No
Usage
f, msg = InitiatePulseTest(tag)
f
msg
tag
2600BS-901-01 Rev. B / May 2013
(on page 3-65)
(on page 7-101)
(on page 3-65)
(on page 5-22)
Affected by
A Boolean flag; this flag is true when the pulse was successfully configured,
false when errors are encountered
A string message; if the f flag is false, msg contains an error message; if it is
true, msg contains a string that indicates successful configuration
Numeric identifier of the pulse configuration to be initiated
Section 7: TSP command reference
Enable high-capacitance mode.
Charge the capacitor.
The parameters passed on to the
i_threshold_measure() function in this
example are:
smu = smua
levelv = 0 V
limiti = 1 A
sourcedelay = 300 ms
measurei = 10 µA range
measuredelay = 100 ms
threshold = 1 µA
timeout = 1 s
The levels and delays will depend on the
value and type of capacitor used.
Sets pass = true if the current is
measured below 1 µA in less than 1 second.