Series 2600B System SourceMeter® Instrument Reference Manual
Model 2611B/2612B/2614B calibration steps
1
Function
Voltage Source
and Measure
Current Source and Measure
1. Calibrate only the source for the CALA sense steps.
2. Steps must be performed in the order shown.
3. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-30. Calibration polarities must also be
set as shown in the procedures.
4. Output must be off before changing to the CALA sense mode.
Model 2634B/2635B/2636B calibration steps
1
Function
Voltage Source
and Measure
Current Source and Measure
1. Calibrate only the source for the CALA sense steps.
2. Steps must be performed in the order shown.
3. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-30. Calibration polarities must also be
set as shown in the procedures.
4. Output must be off before changing to the CALA sense mode.
5. For Current Measure only.
6. This range is only available on the Models 2635B and 2636B.
2600BS-901-01 Rev. B / May 2013
2
Calibration steps
Calibration points
200 mV
±1e-30, ±180 mV
200 mV
±1e-30, ±180 mV
2 V
±1e-30, ±1.8 V
2 V
±1e-30, ±1.8 V
20 V
±1e-30, ±18 V
200 V
±1e-30, ±180 V
100 nA
±1e-30, ±90 nA
1 μA
±1e-30, ±0.9 μA
10 μA
±1e-30, ±9 μA
100 μA
±1e-30, ±90 μA
1 mA
±1e-30, ±0.9 mA
1 mA
±1e-30, ±0.9 mA
10 mA
±1e-30, ±9 mA
100 mA
±1e-30, ±90 mA
1 A
±1e-30, ±0.9 A
1.5 A
±1e-30, ±1.35 A
10 A
±1e-30, ±2.4 A
2
Calibration steps
Calibration points
200 mV
±1e-30, ±180 mV
200 mV
±1e-30, ±180 mV
2 V
±1e-30, ±1.8 V
2 V
±1e-30, ±1.8 V
20 V
±1e-30, ±18 V
200 V
±1e-30, ±180 V
5, 6
±1e-30, ±90 pA
100 pA
±1e-30, ±0.9 nA
1 nA
±1e-30, ±9 nA
10 nA
±1e-30, ±90 nA
100 nA
±1e-30, ±0.9 μA
1 μA
±1e-30, ±9 μA
10 μA
±1e-30, ±90 μA
100 μA
±1e-30, ±0.9 mA
1 mA
±1e-30, ±0.9 mA
1 mA
±1e-30, ±9 mA
10 mA
±1e-30, ±90 mA
100 mA
±1e-30, ±0.9 A
1 A
±1e-30, ±1.35 A
1.5 A
Appendix B: Calibration
3
4
Sense mode
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
3
4
Sense mode
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
B-23