Keithley 2601B Reference Manual page 617

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Usage
operationRegister = status.operation.instrument.digio.trigger_overrun.condition
operationRegister = status.operation.instrument.digio.trigger_overrun.enable
operationRegister = status.operation.instrument.digio.trigger_overrun.event
operationRegister = status.operation.instrument.digio.trigger_overrun.ntr
operationRegister = status.operation.instrument.digio.trigger_overrun.ptr
status.operation.instrument.digio.trigger_overrun.enable = operationRegister
status.operation.instrument.digio.trigger_overrun.ntr = operationRegister
status.operation.instrument.digio.trigger_overrun.ptr = operationRegister
operationRegister
Details
These attributes are used to read or write to the operation status digital I/O overrun registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For example, if a value
of 1.02600e+03 (which is 1026) is read as the value of the condition register, the binary equivalent is
0000 0100 0000 0010. This value indicates that bit B1 and bit B10 are set.
B15
B14
**
>
0
0
* Least significant bit
** Most significant bit
A set bit indicates that the specified digital I/O line generated an action overrun when it was triggered to generate
an output trigger.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
page E-1) and
the following table.
Bit
Value
B0
Not used
B1
status.operation.instrument.digio.trigger_overrun.LINE1
B2
status.operation.instrument.digio.trigger_overrun.LINE2
B3
status.operation.instrument.digio.trigger_overrun.LINE3
B4
status.operation.instrument.digio.trigger_overrun.LINE4
B5
status.operation.instrument.digio.trigger_overrun.LINE5
B6
status.operation.instrument.digio.trigger_overrun.LINE6
B7
status.operation.instrument.digio.trigger_overrun.LINE7
B8
status.operation.instrument.digio.trigger_overrun.LINE8
B9
status.operation.instrument.digio.trigger_overrun.LINE9
B10
status.operation.instrument.digio.trigger_overrun.LINE10
B11
status.operation.instrument.digio.trigger_overrun.LINE11
B12
status.operation.instrument.digio.trigger_overrun.LINE12
B13
status.operation.instrument.digio.trigger_overrun.LINE13
B14
status.operation.instrument.digio.trigger_overrun.LINE14
B15
Not used
2600BS-901-01 Rev. B / May 2013
The status of the operation status digio I/O overrun register; a zero (0) indicates no
bits set (also send 0 to clear all bits); other values indicate various bit settings
B13
B12
B11
B10
>
>
>
>
0
0
0
1
Enable and transition registers
B9
B8
B7
B6
B5
>
>
>
>
0
0
0
0
(on page E-19). The individual bits of this register are defined in
Section 7: TSP command reference
B4
B3
B2
B1
>
>
>
>
>
0
0
0
0
1
Status register set contents
Description
Not applicable
Bit B1 decimal value: 2
Bit B2 decimal value: 4
Bit B3 decimal value: 8
Bit B4 decimal value: 16
Bit B5 decimal value: 32
Bit B6 decimal value: 64
Bit B7 decimal value: 128
Bit B8 decimal value: 256
Bit B9 decimal value: 512
Bit B10 decimal value: 1,024
Bit B11 decimal value: 2,048
Bit B12 decimal value: 4,096
Bit B13 decimal value: 8,192
Bit B14 decimal value: 16,384
Not applicable
B0
*
0
(on
7-277

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents