Trigger.clear(); Trigger.event_Id - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference

trigger.clear()

This function clears the command interface trigger event detector.
Type
TSP-Link accessible
Function
No
Usage
trigger.clear()
Details
The trigger event detector indicates if a trigger event has been detected since the last trigger.wait() call.
trigger.clear() clears the trigger event detector and discards the history of command interface trigger
events.
Also see
trigger.wait()
(on page 7-350)

trigger.EVENT_ID

This constant contains the command interface trigger event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.EVENT_ID
eventID
Details
You can set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
command interface trigger events.
Example
trigger.timer[1].stimulus = trigger.EVENT_ID
Also see
None
7-342
Series 2600B System SourceMeter® Instrument Reference Manual
Affected by
Affected by
The event ID for the command interface triggers
Where saved
Default value
Where saved
Default value
Sets the trigger stimulus of trigger
timer 1 to the command interface
trigger event.
2600BS-901-01 Rev. B / May 2013

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents