Section 7: TSP command reference

status.measurement.instrument.smuX.*

This attribute contains the registers of the measurement event SMU X summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
measurementRegister = status.measurement.instrument.smuX.condition
measurementRegister = status.measurement.instrument.smuX.enable
measurementRegister = status.measurement.instrument.smuX.event
measurementRegister = status.measurement.instrument.smuX.ntr
measurementRegister = status.measurement.instrument.smuX.ptr
status.measurement.instrument.smuX.enable = measurementRegister
status.measurement.instrument.smuX.ntr = measurementRegister
status.measurement.instrument.smuX.ptr = measurementRegister
measurementRegister
X
Details
These attributes are used to read or write to the measurement event SMU X summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For example,
assume the value 257 is returned for the enable register. The binary equivalent is 0000 0001 0000 0001. This
value indicates that bit B0 (VLMT) and bit B8 (BAV) are set.
B15
B14
**
>
0
0
* Least significant bit
** Most significant bit
7-262
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The status of the instrument measurement status SMU X summary register; a
zero (0) indicates no bits set (also send 0 to clear all bits); other values indicate
various bit settings
Source-measure unit (SMU) channel (for example
status.measurement.instrument.smua.enable applies to SMU
channel A)
B13
B12
B11
B10
>
>
>
>
0
0
0
0
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
- -
Not saved
Not saved
Not saved
Not saved
Not saved
B9
B8
B7
B6
B5
>
>
>
>
0
1
0
0
Default value
- -
Not applicable
0
0
0
387 (All bits set)
B4
B3
B2
B1
>
>
>
>
>
0
0
0
0
0
2600BS-901-01 Rev. B / May 2013
B0
*
1