Tsplink.trigger[N].Overrun - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
When programmed to any mode except tsplink.TRIG_BYPASS, the output state of the I/O line is controlled by
the trigger logic, and the user-specified output state of the line is ignored.
When the trigger mode is set to tsplink.TRIG_RISING, the user-specified output state of the line is
examined. If the output state selected when the mode is changed is high, the actual mode used will be
tsplink.TRIG_RISINGA. If the output state selected when the mode is changed is low, the actual mode used
will be tsplink.TRIG_RISINGM.
mode stores the trigger mode as a numeric value when the attribute is read.
To control the line state, use the tsplink.TRIG_BYPASS mode with the tsplink.writebit() and the
tsplink.writeport() commands.
Example
tsplink.trigger[3].mode =
tsplink.TRIG_RISINGM
Also see
digio.writebit()
digio.writeport()
tsplink.trigger[N].assert()
tsplink.trigger[N].clear()

tsplink.trigger[N].overrun

tsplink.trigger[N].release()
tsplink.trigger[N].reset()
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()
tsplink.trigger[N].overrun
This attribute indicates if the event detector ignored an event while in the detected state. This attribute is not
available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
overrun = tsplink.trigger[N].overrun
overrun
N
Details
Indicates that an event was ignored because the event detector was in the detected state when the event was
detected.
Indicates the overrun state of the event detector built into the line itself.
It does not indicate whether an overrun occurred in any other part of the trigger model or in any other detector
that is monitoring the event.
It does not indicate output trigger overrun. Output trigger overrun indications are provided in the status model.
2600BS-901-01 Rev. B / May 2013
(on page 7-58)
(on page 7-59)
(on page 7-356)
(on page 7-356)
(on page 7-359)
(on page 7-360)
(on page 7-361)
(on page 7-362)
(on page 7-364)
Affected by
Instrument reset
Recall setup
TSP-Link trigger N clear
TSP-Link trigger N reset
Trigger overrun state
The trigger line (1 to 3)
Section 7: TSP command reference
Sets the trigger mode for
synchronization line 3 to
tsplink.TRIG_RISINGM.
Where saved
Default value
Not applicable
Not applicable
7-359

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents