Section 7: TSP command reference
Example
trigger.timer[3].delaylist = {50e-6, 100e-6, 150e-6}
DelayList = trigger.timer[3].delaylist
for x = 1, table.getn(DelayList) do
print(DelayList[x])
end
Also see
trigger.timer[N].reset()

trigger.timer[N].EVENT_ID

This constant specifies the trigger timer event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.timer[N].EVENT_ID
eventID
N
Details
This constant is an identification number that identifies events generated by this timer.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to events
from this timer.
Example
trigger.timer[1].stimulus = tsplink.trigger[2].EVENT_ID
Also see
None

trigger.timer[N].overrun

This attribute indicates if an event was ignored because of the event detector state.
Type
TSP-Link accessible
Attribute (R)
Yes
7-346
Series 2600B System SourceMeter® Instrument Reference Manual
(on page 7-348)
Affected by
The trigger event number
Trigger timer number (1 to 8)
Affected by
Instrument reset
Recall setup
Trigger timer N clear
Trigger timer N reset
Set a delay list on trigger timer 3
with three delays (50 µs, 100 µs,
and 150 µs).
Read the delay list on trigger
timer 3.
Output (assuming the delay list was
set to 50 µs, 100 µs, and 150 µs):
5.000000000e-05
1.000000000e-04
1.500000000e-04
Where saved
Default value
Sets the trigger stimulus of
trigger timer 1 to the TSP-
Link trigger 2 event.
Where saved
Default value
Not applicable
false