17-18
SCPI Command Reference
Table 17-7
STATus command summary
Command
:STATus
:MEASurement
[:EVENt]?
:ENABle <NDN>
or <NRf>
:ENABle?
:CONDition?
:OPERation
[:EVENt]?
:ENABle <NDN>
or <NRf>
:ENABle?
:CONDition?
:QUEStionable
[:EVENt]?
:ENABle <NDN>
or <NRf>
:ENABle?
:CONDition?
:PRESet
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
Notes:
1. Commands in this subsystem are not affected by *RST and :SYSTem:PREset. The effects of cycling power, *CLS and
:STATus:PRESet, are explained by the following notes.
2. Event registers — Power-up and *CLS clears all bits. :STATus:PRESet has no effect.
3. Enable registers — Power-up and :STATus:PRESet clears all bits. *CLS has no effect. Accepts the SCPI 1995.0 mandated (non-
decimal numeric) format (#H, #Q, or #B).
4. Error queue — Power-up and *CLS clears all bits of the registers.
5. Error queue messages — Power-up clears list of messages. *CLS and :STATus:PRESet have no effect.
6. Register query commands — The format for the response messages (ASCII, hexadecimal, octal, or binary) depends on which data
format is presently selected. See the :FORMat:SREGister command.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Description
Control measurement event registers:
Read the event register.
Program the enable register.
Read the enable register.
Read the condition register.
Control operation status registers:
Read the event register.
Program the enable register.
Read the enable register.
Read the condition register.
Control questionable status registers:
Read the event register.
Program the enable register.
Read the enable register.
Read the condition register.
Return status registers to default states.
Path to access error queue:
Read the most recent error message.
Specify error and status messages for error queue.
Read the enabled messages.
Specify messages not to be placed in error queue.
Read the disabled messages.
Clears all messages from error queue.
Models 2500 and 2502 User's Manual
6
6
6
6
6
6
6
6
6
Default
parameter SCPI
Note 1
Note 2
Note 3
Note 2
Note 3
Note 2
Note 3
Note 4
Note 5
Note 5

Hide quick links: