Smux.contact.speed - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Details
If you attempt to perform a contact resistance measurement when any of the following conditions exist, an error
will be generated.
When the output is on and any of the following:
SMU is a current source with current range set to less than 1 mA (error code 5065, "I range too low for
contact check")
SMU is a voltage source with current limit set to less than 1 mA (error code 5050, "I limit too low for
contact check")
When the output is off and any of the following:
The output off mode is High-Z (error code 5048, "Contact check not valid with HIGH-Z OUTPUT off")
The output off mode is Normal with the smuX.source.offfunc attribute set to
smuX.OUTPUT_DCVOLTS and the off current limit set to less than 1 mA (error code 5066,
"source.offlimiti too low for contact check")
The output off mode is Normal with the smuX.source.offfunc attribute set to
smuX.OUTPUT_DCAMPS and the source range is less than 1 mA (error code 5065, "I range too low for
contact check")
Example
if not smua.contact.check() then
smua.contact.speed = smua.CONTACT_SLOW
rhi, rlo = smua.contact.r()
print(rhi, rlo)
exit()
end
Also see
Contact check connections
Contact check measurements
smuX.contact.check()

smuX.contact.speed

smuX.contact.speed
This attribute stores the speed setting for contact check measurements. This attribute is not available on the
Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
speedSetting = smuX.contact.speed
smuX.contact.speed = speedSetting
speedSetting
X
2600BS-901-01 Rev. B / May 2013
(on page 2-54)
(on page 2-44)
(on page 7-193)
(on page 7-195)
Affected by
Instrument reset
SMU reset
Recall setup
The speed setting. Set to one of the following:
0 or smuX.CONTACT_FAST
1 or smuX.CONTACT_MEDIUM
2 or smuX.CONTACT_SLOW
Source-measure unit (SMU) channel (for example, smua.contact.speed applies
to SMU channel A)
Section 7: TSP command reference
Check contacts against threshold.
Set speed for SMU channel A to slow.
Get resistance readings.
Output contact resistances to the host.
Terminate execution.
Where saved
Default value
Not saved
0 (smuX.CONTACT_FAST)
7-195