Section 7: TSP command reference
Also see
Operation Status Registers

status.operation.calibrating.*

This attribute contains the operation status calibration summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
operationRegister = status.operation.calibrating.condition
operationRegister = status.operation.calibrating.enable
operationRegister = status.operation.calibrating.event
operationRegister = status.operation.calibrating.ntr
operationRegister = status.operation.calibrating.ptr
status.operation.calibrating.enable = operationRegister
status.operation.calibrating.ntr = operationRegister
status.operation.calibrating.ptr = operationRegister
operationRegister
Details
These attributes are used to read or write to the operation status calibration summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
page E-1) and
the following table.
Bit
Value
B0
Not used
B1
status.operation.calibrating.SMUA
B2
status.operation.calibrating.SMUB
B3-B15
Not used
7-272
(on page E-9)
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The status of the operation calibrating event register; a zero (0) indicates no bits set
(also send 0 to clear all bits); other values indicate various bit settings
Enable and transition registers
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
(All bits set)
(on page E-19). The individual bits of this register are defined in
Description
Not applicable.
Set bit indicates that SMU A is unlocked for
calibration.
Bit B1 decimal value: 2
Binary value: 0000 0010
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set bit
indicates that SMU B is unlocked for calibration.
Bit B2 decimal value: 4
Binary value: 0000 0100
Not applicable.