Section 2: General operation
Step 5: Observe readings on the display.
Press the TRIG key if necessary to trigger the instrument to begin taking readings. The readings are
on the top line, and source and limit values are on the bottom line.
Step 6: Turn the output off
When finished, turn the output off by pressing the OUTPUT ON/OFF control. The OUTPUT indicator
light switches off.
Remote source-measure commands
Basic source-measurement procedures can also be performed through a remote interface. To do this,
send the appropriate commands. The following table summarizes basic source-measure commands.
See
Introduction to TSP operation
Basic source-measure commands
Command*
smuX.measure.autorangei = smuX.AUTORANGE_ON
smuX.measure.autorangev = smuX.AUTORANGE_ON
smuX.measure.autorangei = smuX.AUTORANGE_OFF
smuX.measure.autorangev = smuX.AUTORANGE_OFF
smuX.measure.rangei = rangeval
smuX.measure.rangev = rangeval
reading = smuX.measure.i()
reading = smuX.measure.v()
iReading, vReading = smuX.measure.iv()
reading = smuX.measure.r()
reading = smuX.measure.p()
smuX.source.autorangei = smuX.AUTORANGE_ON
smuX.source.autorangev = smuX.AUTORANGE_ON
smuX.source.autorangei = smuX.AUTORANGE_OFF
smuX.source.autorangev = smuX.AUTORANGE_OFF
smuX.source.func = smuX.OUTPUT_DCVOLTS
smuX.source.func = smuX.OUTPUT_DCAMPS
smuX.source.leveli = sourceval
smuX.source.levelv = sourceval
smuX.source.limiti = level
smuX.source.limitv = level
smuX.source.limitp = level
smuX.source.output = smuX.OUTPUT_ON
smuX.source.output = smuX.OUTPUT_OFF
smuX.source.rangei = rangeval
smuX.source.rangev = rangeval
smuX.sense = smuX.SENSE_LOCAL
smuX.sense = smuX.SENSE_REMOTE
* smuX: For Models 2601B, 2611B, and 2635B, this value is smua (SMU Channel A); for Models 2602B, 2604B, 2612B, 2614B,
2634B, and 2636B, this value can be smua (for SMU Channel A) or smub (for SMU Channel B).
2-34
Series 2600B System SourceMeter® Instrument Reference Manual
(on page 5-1) for more information on using these commands.
Description
Enable current measure autorange.
Enable voltage measure autorange.
Disable current measure autorange.
Disable voltage measure autorange.
Set current measure range.
Set voltage measure range.
Request a current reading.
Request a voltage reading.
Request a current and voltage reading.
Request a resistance reading.
Request a power reading.
Enable current source autorange.
Enable voltage source autorange.
Disable current source autorange.
Disable voltage source autorange.
Select voltage source function.
Select current source function.
Set current source value.
Set voltage source value.
Set current limit.
Set voltage limit.
Set power limit.
Turn on source output.
Turn off source output.
Set current source range.
Set voltage source range.
Select local sense (2-wire).
Select remote sense (4-wire).