The TEST system contains the commands for performing selftest routines, and for
direct adjustment of the hardware assemblies (:TEST:DIRect).
The self tests return a "0" if the test is performed successfully, otherwise a value other
than "0" is returned. None of the commands of this system have an *RST value.
Improper use can destroy the assembly
The respective hardware assembly responds directly to the :TEST:DIRect command;
any safety mechanisms are bypassed. The command is intended for servicing purpo-
ses and should be used only by the Rohde & Schwarz service personnel.
:TEST<hw>:ALL:STARt
:TEST<hw>:ALL:RESult?
:TEST<hw>:DIRect
:TEST<hw>:ALL:STARt
Starts a self-test on all installed hardware options.
To query the result, use the command
Example:
Usage:
Manual operation:
:TEST<hw>:ALL:RESult?
Queries the result of the performed self-test (command
on page 448).
Operating Manual 1407.0806.32 ─ 21
integer
Wait time in ms
Range:
0 to 10000
*RST:
0
:SYSTem:WAIT 10000
// waits 10s before resetting the instrument
*RST
Setting only
.................................................................................................. 448
................................................................................................448
........................................................................................................ 449
:TEST<hw>:ALL:RESult?
See
:TEST<hw>:ALL:RESult?
Event
See
"Start Selftest"
Remote Control Commands
on page 448
on page 101
:TEST<hw>:ALL:STARt