R&S SMB100A Operating Manual page 410

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Example:
Usage:
Options:
Manual operation:
[:SOURce<hw>]:PULM:TRAin:OFFTime <OffTime>
Fills the Off-time part of the selected file with data.
*RST does not affect data lists.
Parameters:
<OffTime>
Example:
Options:
Manual operation:
[:SOURce<hw>]:PULM:TRAin:OFFTime:POINts?
Queries the length (in points) of the off-time component of the selected list.
Return values:
<Points>
Operating Manual 1407.0806.32 ─ 21
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:DEL 'P_FIVE'
deletes the list P_FIVE
Setting only
R&S SMB-K27 (Pulse Train)
See
"Pulse Train Data – Pulse Generator"
Offtime#1{, Offtime#2, ...} | binary block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data.
When block data is transferred, 8 (4) bytes are always interpre-
ted as a floating-point number with double accuracy (see the
command FORMat:DATA).
The maximum length is 2047 values.
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:SEL 'P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:OFFT 10ns,30ns,40ns,...
specifies the off-time values in P_INCR. If the list already con-
tains data, it is overwritten.
R&S SMB-K27 (Pulse Train)
See
"Edit Pulse Train Data"
integer
Range:
0 to 2047
*RST:
0
Remote Control Commands