The "Remote Channel Settings" dialog contains the GPIB address and displays the
VISA resource strings provided for remote control via the various interfaces.
GPIB channel address
Sets the address of the GPIB channel the instrument is connected to.
Remote command:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
RS232 using USB adapter
Remote control via a serial interface is possible via a USB. The controller and the
instrument must be connected with the external USB/serial-adapter R&S TS1-USB
(see recommended extras in the data sheet) and a serial crossover (null modem)
cable. A USB connection requires the VISA library to be installed on the controller.
VISA will detect and configure the R&S SMB automatically when the USB connection
is established.
In addtion, you can also use a Bluetooth connection for remote control via the serial
interface. The settings are effective for both interfaces (see also
"Security",
Baud Rate ← RS232 using USB adapter
Sets the baudrate for the serial remote control interface.
Remote command:
:SYSTem:COMMunicate:SERial:BAUD
Parity ← RS232 using USB adapter
Sets the parity for the serial remote control interface.
Remote command:
:SYSTem:COMMunicate:SERial:PARity