Save/Recall; Factory Preset - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Via remote control, there is no password required.
Remote command:
:SYSTem:ULOCk
:SYSTem:DLOCk
:SYSTem:KLOCk
Security Password
Enters the password that is required to enable or to disable the settings protected by a
security password. Default is '123456'.
Note: It is highly recommended that you to change the default security password
before connecting the instrument to the network.
All settings are only accepted after the "Accept" button is pressed.
Accept
Applies the modified settings, provided the security password is entered correctly.
Note: This action can not be undone. Keep in mind, that a changed password is never
reset, even if you perform a factory preset.
Bluetooth Pin
Sets the Bluetooth pin of an external Bluetooth device. The pin is required to enable
remote control via an external Bluetooth device.
Requires a Bluetooth adapter (recommended extra, see data sheet) .
5.2.3.19

Save/Recall

The "Save/Recall" submenu can also be called up with the FILE key and is therefore
described in the section of this key (see
ment Data - File
5.2.3.20

Factory Preset

The "Factory Preset" dialog provides a function to reset the instrument's settings to
their factory state. This function is activated by pressing the "Execute Factory Preset"
button.
Factory Preset
Reset the instrument's settings to their factory state.
Note: "Factory Preset" resets the "Remote Channel" and network settings to the
default values.
Executing "Factory Preset" via remote control terminates the connection to the instru-
ment, if these settings had been configured to values different to the default ones.
Operating Manual 1407.0806.32 ─ 21
on page 437
on page 436
on page 436
Key",
on page 126).
General Instrument Settings
Chapter 5.2.8, "Storing and Loading Instru-
Instrument Function
120