R&S SMB100A Operating Manual page 398

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Parameters:
<Mode>
Example:
Manual operation:
[:SOURce<hw>]:POWer:POWer <Power>
Sets the RF level of the RF output connector.
The level entered with this command corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note: The SCPI command
AMPLitude]
Parameters:
<Power>
Example:
Manual operation:
[:SOURce<hw>]:POWer:SPC:CRANge <PowCntrlCRange>
Defines the capture range of the power control system.
Within the range:
Target Level +/- Catch Range
Operating Manual 1407.0806.32 ─ 21
CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW and FIXed are synonyms. To set the output level value, use
the command
IMMediate][:AMPLitude].
SWEep
Operates in power sweep mode.
Set the range and current level with the commands
SOURce<hw>]:POWer:STARt,
and [:SOURce<hw>]:POWer:MANual.
STOP
*RST:
CW
POW:MODE SWEep
selects the SWEep mode using the
POW:STAR; POW:STOP; POW:MAN settings.
See
"State - Level Sweep"
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
sets the level of the "Level" display, i.e. the level containing offset.
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The values are given in the data sheet.
Increment: 0.01
*RST:
-30
SOUR:POW:POW 15
sets the RF level at output to 15 dBm.
See
"Amplitude"
Remote Control Commands
[:SOURce<hw>]:POWer[:LEVel][:
[:SOURce<hw>]:POWer:
on page 182
on page 147
SOURce Subsystem
[:
383