Unit Subsystem; Trigger<Hw>[:Sweep][:Immediate]; Trigger<Hw>[:Immediate] - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Manual operation:
:TRIGger<hw>[:SWEep][:IMMediate]
Starts all sweeps which are activated for the respective path. The command starts all
sweeps which are activated.
The sweep to be executed depends on the respective MODE setting
(:SOUR:SWEep:POW|FREQ:MODE and :SOUR:LFO:SWEep[:FREQ]:MODE).
The command corresponds to the manual-control command "Execute Trigger".
Example:
Usage:
Manual operation:
:TRIGger<hw>[:IMMediate]
The command immediately starts the activated sweep.
The command performs a single sweep and therefore applies to sweep mode AUTO
with sweep source SINGle. Use the commands
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING, and SOUR:SWE:FREQ|POW:MODE, or
SOUR:LFO:SWE:[FREQ:]MODE to set the respective sinlge sweep. You can alterna-
tively use an IMMediate command instead of the respective SWEep:
[FREQ:]|POW:EXECute command.
Example:
Usage:
Manual operation:

7.18 UNIT Subsystem

The UNIT subsystem contains the commands specifying which units are valid if no unit
is indicated in a command. These settings are valid for the entire instrument.
:UNIT:ANGLe <Angle>
Defines the default unit for the phase modulation angle. It is not valid for commands
which determine angle values, e.g. RF phase. It does not influence the manual control
parameter unit and the display.
Parameters:
<Angle>
Operating Manual 1407.0806.32 ─ 21
See
"Mode"
on page 221
TRIG
starts all active sweeps.
Event
See
"Execute Single Sweep - Frequency Sweep"
TRIG
starts all active sweeps.
Event
See
"Execute Single Sweep"
DEGRee | RADian
*RST:
RADian
Remote Control Commands
UNIT Subsystem
on page 178
on page 223