Remote command:
[:SOURce<hw>]:POWer:MODE
Mode - Level Sweep
Selects the level sweep instrument operating mode and the sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
"Single"
Operating Manual 1407.0806.32 ─ 21
on page 382
Sets an automatically repeated sweep cycle.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR AUTO
SOUR:POW:MODE SWE
Sets a single sweep cycle. The sweep is triggered by the "Execute
Single Sweep" button, or by means remote trigger commands, e.g.
*TRG.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:POW:EXEC