Parameters:
<Start>
Example:
Manual operation:
[:SOURce<hw>]:POWer:STEP[:INCRement] <Increment>
Sets the step width for
To adjust the level step by step with this step size, use the POW:UP and POW:UP com-
mands.
Note: This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the level accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
Example:
Manual operation:
[:SOURce<hw>]:POWer:STEP:MODE <Mode>
Activates (USER) or deactivates (DECimal) the user-defined step width used when
varying the level value with the level values UP/DOWN. The command is linked to set-
ting "Variation Active" for manual control, i.e. the command also activates/deactivates
the user-defined step width used when varying the level value with the rotary knob.
Parameters:
<Mode>
Operating Manual 1407.0806.32 ─ 21
float
Determines the first level value of the sweep setting range.
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
-30
POW:STAR -20 dBm
sets the start level for the level sweep to -15 dBm for RF output
A.
See
"Start Level - Level Sweep"
POW:STEP:MODE
USER.
float
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
1
POW:STEP 2
sets the step width for entering the RF level to 2 dB.
See
"Variation Step"
on page 150
DECimal | USER
*RST:
DECimal
Remote Control Commands
SOURce Subsystem
on page 185
386