R&S SMB100A Operating Manual page 501

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Error
Code
50
140
180
182
183
200
201
202
203
204
241
242
Operating Manual 1407.0806.32 ─ 21
Status Information, Error Messages and Troubleshooting
Error
Description
Extern reference out of
External reference is selected but no
range or disconnected
external signal is applied or the signal
is out of range.
This modulation forces
A modulation has been switched on
other modulations off
which cannot be used at the same
time as an already active modulation.
The previous modulation has been
switched off.
Example: Enabling FM modulation
switches PM modulation off.
Adjustment failed
Adjustment could not be executed
Adjustment data miss-
Adjustment data are missing.
ing
Adjustment data inva-
Adjustment data are invalid and must
lid
be restored.
Cannot access hard-
The data transmission to a module
ware
was unsuccessful.
Hardware revision out
A later version of certain parts of the
of date
instrument is necessary to execute
the function selected.
Cannot access the
A error occurs when writing or reading
EEPROM
a EEPROM.
Invalid EEPROM data
Reading a EEPROM is possible, how-
ever the data are inconsistent.
Driver initialization
Initialization of a driver fails when
failed
booting the instrument firmware.
No current list
There is no list selected. To execute
the required operation, a list has to be
selected in the related menu.
Unknown list type
The list type selected is not valid for
specified
the required operation.
For instance, the file extension for
waveform list files is *.wv. It is not
possible to enter another file exten-
sion when selecting a list.
Device-Specific Error Messages
Remedy
Check the selected refer-
ence signal source (internal
or external) in the "Setup >
Reference Oscillator" dialog.
Change setting to 'internal' if
no appropriate external
source is available.
The adjustment data have to be
generated first by an internal or
external adjustment or to be loa-
ded into the device (see
ter 5.2.3.1, "Internal Adjust-
ments",
on page 92.
The adjustment data have to be
generated first by an internal or
external adjustment or to be loa-
ded into the instrument.
The adjustment data have to be
generated again by an internal or
external adjustment or to be loa-
ded into the instrument.
The module is not installed, not
properly installed or missing.
The driver does not support the
installed version of a module.
The EEPROM might be defect
and has to be replaced.
The driver is not compatible with
the hardware or software config-
uration of the instrument.
If no list is available, a new list
must be created.
Check the selected list type.
Chap-
486

Advertisement