R&S
SMB100A
To indicate the active state, the RF block is displayed in blue. An RF signal with the
default frequency and level settings is output, i.e. 1 GHz and -30 dBm.
The sweep is not yet active, it must be triggered in the sweep dialog.
4. Trigger RF frequency sweep
a) Press the WINBAR key to switch to the "RF Frequency Sweep" dialog.
A linear single sweep signal is now present at the RF output, starting at 100 MHz.
The sweep is processed in 1 MHz steps with dwell time of 10 ms per step up to the
stop frequency of 500 MHz.
The sweep starts at 100 MHz, stops at 500 MHz in 1 MHz steps is output with a
dwell time of 10 ms per step.
Operating Manual 1407.0806.32 ─ 21
Turn the rotary knob to select the "Execute Single Sweep" button.
Press the rotary knob to trigger (start) the frequency sweep.
Getting Started
Example of Setup
63