[:Source<Hw>]:Sweep:power:spacing:mode; [:Source<Hw>]:Sweep:power:step[:Logarithmic] - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Manual operation:
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the sweep spacing mode. The sweep spacing for level sweeps is always lin-
ear.
Return values:
<Mode>
Example:
Usage:
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] <Logarithmic>
Sets a logarithmically determined sweep step size for the RF level sweep. It is
expressed in decibels and you must enter the value and the unit dB with the command.
The level is increased by a logarithmically calculated fraction of the current level
according to:
step_size
Level
with Level
sweep steps
This parameter correlates with the number of steps
range as follows:
level_points = ((Level
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
Parameters:
<Logarithmic>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
See
"Shape - RF Level Sweep"
LINear
*RST:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result: LIN
linear spacing
Query only
= Level
+ step_size
n+1
n
= Level
+ step_size
n+1
n
< level
, step_size = SWE:POW:STEP[:LOG] and n = number of
STARt
STOP
- Level
STOP
float
Increment: 0.01
*RST:
SWE:POW:STEP 10dB
sets the step width for logarithmic sweep spacing to 10 dB of the
previous level in each instance (for a level sweep).
See
"Step - Level Sweep"
LINear
x Level
n
n
n+1
) / step_size) + 1)
STARt
1
on page 187
Remote Control Commands
SOURce Subsystem
on page 186