R&S SMB100A Operating Manual page 265

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Note: If the connection cannot be set up, R&S VISA displays an error in the log
view. For information on how to proceed when network failures occur, see
ter 9.5, "Resolving Network Connection
For further information on the functions to read and write to an open session, as
well as the utility applications the software provides, see the R&S VISA User Man-
ual.
Starting a remote control over LAN (using VXI-11)
To set the instrument to remote control, you can use the addressed command &GTR, or
send any command from the controller.
1. Start the R&S VISA Tester and establish the connection to the R&S SMB, see
"Configuring the controller"
2. In the R&S VISA "Basics" tab, enter a SCPI command, e.g. "*IDN?" and confirm
with "Query".
The instrument is switched to remote control when it receives a command from the
controller.
3. Select "Read" to obtain the instrument response.
Tip: If the "Show Log" checkbox is checked R&S VISA displays each VISA function
call in the log-view on the left. If you check the "Write Log" checkbox the log-view
entry is written to the log file as well. You can operate the log-view in two modes:
the "Live Mode" shows only the most recent messages whereas the "View Mode"
allows you to scroll the history.
4. To set, e.g. the frequency, enter SOUR1:FREQ 4 GHz and select "Write".
To check the performed setting, SOUR1:FREQ? and select "Read".
Operating Manual 1407.0806.32 ─ 21
Starting a Remote Control Session
Failures",
on page 247.
Remote Control Basics
on page 487.
Chap-
250