Source:phase Subsystem - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
[:SOURce<hw>]:PGENerator:STATe <State>
Activates/deactivates the output of the video/sync signal at the PULSE VIDEO connec-
tor at the rear of the instrument.
The signal output and the pulse generator are automatically switched on with activation
of pulse modulation if pulse generator is selected as modulation source. The signal
output can be switched off subsequently.
Parameters:
<State>
Example:
Manual operation:

7.13.10 SOURce:PHASe Subsystem

This subsystem contains the commands for adjusting the phase of the RF output signal
relative to a reference signal of the same frequency.
[:SOURce<hw>]:PHASe
[:SOURce<hw>]:PHASe:REFerence
[:SOURce<hw>]:PHASe <Phase>
Sets the phase variation relative to the current phase. The variation is specified in
RADians.
Parameters:
<Phase>
Example:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
0 | 1 | OFF | ON
*RST:
0
PULM:SOUR INT
selects the internal pulse generator as source for pulse modula-
tion
PULM:STAT ON
activates pulse modulation. The pulse generator and the output
of the signals at the PULSE VIDEO connector are automatically
activated in addition.
PGEN:STAT OFF
deactivates the output of the pulse signal by the pulse generator
at the PULSE VIDEO connector. The pulse modulation of the RF
carrier must be activated with command SOURce:PULM:STATe.
See
"Video Sync Signal State - Pulse Generator"
................................................................................................. 372
................................................................................ 373
float
Range:
-720 to 720
Increment: 0.1
*RST:
0
PHAS 0.1 RAD
changes the phase by 0.1 RAD relative to the current phase.
PHAS:REF
adopts the set phase as the current phase.
See
"Delta Phase"
on page 141
Remote Control Commands
SOURce Subsystem
on page 227
372