SMB100A
State
Activates the LF frequency sweep signal generation.
Note:
Activating a sweep mode automatically deactivates other sweeps and the list mode.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE
Mode
Selects the LF frequency sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Operating Manual 1407.0806.32 ─ 21
Generates a continuously repeating sweep signal immediately after
activating the sweep mode.
The sweep steps are performed atomatically, controlled by the dwell
time, see
"Dwell Time - LF Sweep"
Example:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR AUTO
SOUR:LFO:FREQ:MODE SWE
Instrument Function
Modulation Generator and LF Output
on page 350
on page 225.