R&S SMB100A Operating Manual page 184

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
Remote command:
SENSe<ch>[:POWer]:STATus[:DEVice]?
Type
Indicates the type and the serial number of the connected R&S NRP-Z power sensor.
The sensor type is automatically detected.
Remote command:
SENSe<ch>[:POWer]:TYPE?
SENSe<ch>[:POWer]:SNUMber?
State
Activates/deactivates level measurement by the power sensor.
The local state is set with the INIT command. Switching the local state off enhances
the measurement performance.
In remote control, the sensors are set up using the SENSe commands. The remote
measurement is triggered by the READ query which also provides the measurement
results. The state is not influenced by these commands, measurements results can be
retrieved with local State on or off.
The sensor is selected by suffix 1, 2, 3 or 4 in key word SENSe or READ of the com-
mand header.
Suffix 1 denotes the sensor connected at the first USB interface, and suffix 2, 3 and 4
are assigned to further sensors connected via USB. The suffix is identical to the index
which is assigned automatically to each sensor upon connection.
To query the availability of a sensor at a given connector, use the command
SENSe<ch>[:POWer]:STATus[:DEVice]?
Remote command:
:INITiate<ch>[:POWer]:CONTinuous
Level (Peak)
With certain power sensors only, for example R&S NRP-Z81.
Indicates the measured peak level value with the selected unit.
Remote command:
:READ<ch>[:POWer]?
Level (Avg.)
Indicates the measured level value with the selected unit.
Remote command:
:READ<ch>[:POWer]?
Unit
Selects the unit used for result display.
The power sensor provides the measured value in Watt.
In which unit the measured value is indicated is selected here and might be Watt, dBm
or dBuV.
Operating Manual 1407.0806.32 ─ 21
on page 323
on page 324
on page 322
on page 323.
on page 316
on page 317
on page 317
Instrument Function
RF Block
169

Advertisement

Table of Contents
loading

Table of Contents