Performing the Front Panel Tests
To perform the front panel test, you operate the keys at the front panel, and check the
response of the instrument in the "Check Front Panel" dialog. To perform this test
properly, it is essential that you check each key of the front panel. The test is only com-
pleted, when you have veryfied all keys.
During the test, the actual functions of the keys are disabled.
Proceed as follows:
1. Press the SETUP key.
2. Select "Test > Check Front Panel"
The "Check Front Panel" dialog opens.
3. Press a key on the front panel.
Check if the corresponding key in the "Check Front Panel" dialog turns green.
4. Press the same key a second time.
Check that the key in the dialog turns red.
Note: Pressing the same key again has no further effect, with the exception of the
ESC key, see
5. Continue with the next key on the front panel and repeat
keys are tested.
The test is completed, when each key is verified successfully, confirmed by a "Test
passed" message.
Select "OK" to exit the test.
► Press the ESC key a third time.
Exits the "Check Front Panel" dialog, even if you have not yet checked all the keys.
Expected responses:
Pressing a key once (green), pressing twice (red)
Pressing the ESC key a third time exits the dialog.
If you detect a malfunction, for example, you press the front panel key the first time,
and the color of the button in the dialog turns red (instead of green), the front panel key
may be stuck. In this case, contact the Rohde & Schwarz Customer Support Center for
technical support, see http://www.customersupport@rohde-schwarz.com.
5.2.3.10

Shutting Down and Rebooting the Instrument

The POWER ON/STANDBY front panel key switches the instrument from the standby
to the ready state or vice versa. In remote operation form a remote computer or in
Operating Manual 1407.0806.32 ─ 21
Press the ESC key a third
Instrument Function
General Instrument Settings
time.
step 3
to
step 5
until all
102

Advertisement

Table of Contents
loading

Table of Contents