R&S
SMB100A
Risk of injury if feet are folded out
The feet may fold in if they are not folded out completely or if the instrument is shifted.
This may cause damage or injury.
Fold the feet completely in or completely out to ensure stability of the instrument.
Never shift the instrument when the feet are folded out.
When the feet are folded out, do not work under the instrument or place anything
underneath.
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 500 N.
Risk of injury and instrument damage if stacking instruments
A stack of instruments may tilt over and cause injury. Furthermore, the instruments at
the bottom of the stack may be damaged due to the load imposed by the instruments
on top.
Observe the following instructions when stacking instruments:
Never stack more than three instruments with the same dimensions (width and
length). If you need to stack more than three instruments, install them in a rack.
The overall load imposed on the lowest instrument must not exceed 500 N.
All instruments should have the same dimensions (width and length).
If you need to stack smaller instruments on the top, the overall load imposed on the
lowest instrument must not exceed 250 N.
If the instruments have foldable feet, fold them in completely.
Operating Manual 1407.0806.32 ─ 21
Preparing for Use
Putting into Operation
27