*PRE <Value>
Parallel poll register enable
Sets parallel poll enable register to the indicated value. The query returns the contents
of the parallel poll enable register in decimal form.
Parameters:
<Value>
*PSC <Action>
Power on status clear
Determines whether the contents of the ENABle registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action>
*RCL <Number>
Recall
Loads the instrument settings from an intermediate memory identified by the specified
number. The instrument settings can be stored to this memory using the command
*SAV
It also activates the instrument settings which are stored in a file and loaded using the
MMEMory:LOAD <number>, <file_name.extension> command.
Manual operation:

*RST

Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to SYSTem:PRESet.
Usage:
Manual operation:
Operating Manual 1407.0806.32 ─ 21
See
"Hardware Options / Software Options"
Range:
0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
with the associated number.
See
"Recall"
Setting only
See
"Preset"
Remote Control Commands
0 to 255
on page 129
on page 91
Common Commands