[:Source<Hw>]:Sweep[:Frequency]:Execute; [:Source<Hw>]:Sweep[:Frequency]:Lfconnector; [:Source<Hw>]:Sweep[:Frequency]:Mode - R&S SMB100A Operating Manual

Rf and microwave signal generator
Hide thumbs Also See for SMB100A:
Table of Contents

Advertisement

®
R&S
SMB100A
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Starts an RF frequency sweep cycle manually.
The command is only effective in single mode.
Example:
Usage:
Manual operation:
[:SOURce<hw>]:SWEep[:FREQuency]:LFConnector <LfConnector>
Activates the output of a sweep voltage ramp at the LF connector.
The voltage range is set with commands
SOURce:SWEep:FREQuency:OVOLtage:STARt and ...:STOP
Parameters:
<LfConnector>
Example:
Manual operation:
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
Sets the sweep mode.
The keyword [:FREQuency] can be omitted (see example). The command is then
SCPI-compliant.
Operating Manual 1407.0806.32 ─ 21
TRIG:FSW:SOUR SING
SOUR:SWE:FREQ:MODE AUT
SWE:FREQ:EXEC
triggers a frequency sweep at the RF output.
Event
See
"Execute Single Sweep - Frequency Sweep"
0 | 1 | OFF | ON
*RST:
0
SWE:LFC ON
activates the output of a linear voltage ramp from sweep start to
sweep stop at the LF connector.
SWE:OVOL:STAR 0V
SWE:OVOL:STOP 3V
'the voltage at sweep start is 0 Volt and at sweep stop 3 V.
See
"Use LF connector to output sweep voltage - RF Frequency
Sweep"
on page 181
Remote Control Commands
SOURce Subsystem
on page 178
419

Advertisement

Table of Contents
loading

Table of Contents